Serves for measurement of insoluble particles in injections, for example at pharmaceutical plants.
| Optical system | Light-obscuration method |
|---|---|
| Light source | Laser diode (rated output 4.5 mW, wavelength 790 nm) |
| Materials of parts exposed to sample fluid | Flow cell:Synthetic quartz Syringe:Borosilicate glass, PTFE Syringe pump:Kel-F (PCTFE), PTFE Tube, packing, joint:PFA, PTFE, PCTFE, Perflo (special fluorine rubber) |
| Measurement size range | 1 to 20 ranges from 1.3 μm to 100 μm (in 0.1 μm steps) |
| Flow rate | 25 mL syringe:25 mL/min 10 mL syringe:10 mL/min*Factory option |
| Maximum particle number concentration | 10,000 particles/mL(10 μm vicinity PSL particles, coincidence loss max. 10%) |
| Power requirements | 100 V to 240 V AC, 50/60 Hz, approx. 80 VA |
| Dimensions | Approx. 366 (H) x 360 (W) x 236 (D) mm |
| Weight | Approx. 10 kg |